Please enter keyword
Please enter keyword
Do you want to exit?
Yes
Cancel

Micro-Rx Spectrum Analyzer Module

Device:Test Instrument

Vendor: Transcom Instruments

Overview

1.Description of the product Micro-Rx designed with a small shape which allow it to be easily integrate to any instrument. With excellent testing performance and measurement sensitiveness, Micro-Rx suits the testing requirement of the majority of RF signals. Micro- Rx satisfy the needs of general spectrum test, and further secondary development is also allowed based on the API function library. 2.Differentiation and advantage -Frequency range: 10MHz to 4200MHz -DANL: -168 dBm @1GHz (Sensitivity set to High, normalized to 1Hz) -Resolution bandwidth: 10Hz to 5MHz -Signal storage depth of 1Gbit for signal capture and analysis -Small (193mm×93mm×34mm), lightweight (only 0.8kg), and easy to carry -Provide API function library to support secondary development 3.Target market Research and Education, Manufacturer, Operator and System Integrator

Parameter

Function

Sensitivity

Low, Medium and High

Frequency

Frequency range

10MHz to 4.2GHz

Frequency reference

Aging rate: ±1ppm

Frequency readout accuracy:

± ((Center Frequency + 1GHz) x Frequency Reference + Frequency Span Accuracy x Span)

Frequency span accuracy

±1%

Frequency span accuracy

1.1ms to 1600s

2.69ms to 1600s, zero span

Resolution bandwidth

RBW range

10Hz to 5MHz, (1-2-3-5-10 Sequence)

RBW accuracy

RBW1MHz, ±10%

RBW1MHz, ±2%

Amplitude

Measuring range

Display average noise level to +20dBm

Input Attenuator Range

0 to 30dB, 1dB Step

Maximum safe input level

Sensitivity: +30dBm (Low)

Sensitivity: 0dBm (Medium)

Sensitivity: -20dBm (High)

Reference level range

-140 dBm to +20dBm

-190dBm to +70dBm (Ref level offset: ON)

Amplitude accuracy

ATT set to 0 dB, input signal: -5 to -30 dBm; detector set to Positive, Sensitivity set to Low; RBW auto-coupled, all other settings auto-coupled, 23±5

Half hour warm-up required.

±1.5dB

RBW Switching uncertainty

±0.3dB

Input Attenuator uncertainty

±0.6dB

Accuracy of reference level

Reference level: -60dBm, ±0.8dB

Display average noise level (DANL) @1GHz

Input Terminated, Detector set to Positive, Trace Average set to 1000,Span set to 50kHz,Ref set to -100dBm,all other settings auto-coupled, 23±5 .

Normalized to 1 Hz RBW

Sensitivity: Low, -131dBm/Hz (typically

-133dBm/Hz)

Sensitivity: Medium, -151dBm/Hz (typically -153dBm/Hz)

Sensitivity: High, -168dBm/Hz (typically

-169dBm/Hz)

Residual response

-75dBm

Input-related response

l 10M to 1.285G, 3.22G to 3.7G <-70dBc

l 1.625G to 1.775G <-55dBc

l 1.285G to 1.625G,1.775G to 2.35G, 2.71G to 3.22G <-42dBc

l 3.7G to 4.2G <-35dBc

l 2.35G to 2.71G <-25dBc

Second harmonic distortion

1.6GHz to 70dBc

Third-order intercept (TOI)

-10dBm tones, 1MHz apart, Sensitivity set to low, Ref set to -10 dBm

+15dBm

P1dB

+5dBm (nominal)

Phase noise @1GHz

-96dBc/Hz, @10kHz (typically -98dBc/Hz)

-118dBc/Hz, @1MHz (typically -120dBc/Hz)

Storage

Maximum storage depth

1Gbit

Data format

I/Q two-way, 16bit

General

OS

Windows Xp, Windows 7, Windows 8, and Windows 10

Connectors

RF input: N-type, female, 50Ω

USB: USB type C

Power interface: DC12V

Operating environment

Operating temperature: 0°C to 50°C

Storage temperature: -20°C to 70°C

Dimension

193mm×93mm×34mm

Weight

0.8kg

Technical specifications
This technical specifications include the influence of probability distribution, measurement uncertainty and environmental factors on the instrument performance. It guarantees the performance under the following conditions.
• The instrument is ON and warmed up for 30min.
• The instrument internal reference signal is applied.
Testing temperature is 23±5 ºC, unless other specific condition applied.
Typical value
Additional description does not cover all performance information of the product guarantee. Unless otherwise specified, the typical value refers to the indicator or technical specification with which more than 80% of products comply under 23 ± 5 ºC. The measurement uncertainty is excluded. Micro-Rx should be within the calibration period.
Nominal value
The nominal value refers to the characteristic description or design range. It is not tested or covered by the product. Micro-Rx should be within the calibration period.