Please enter keyword
Please enter keyword
Do you want to exit?
Yes
Cancel

SpecMini Handheld Spectrum Analyzer

Device:Test Instrument

Vendor: Transcom Instruments

Overview

1.Description of the product SpecMini is the first Android hand-held spectrum analyzer. It features high testing sensitivity, light weight, compact size and portable design. Android operating system and highresolution touch screen allow testing and measurement can be performed user-friendly. With excellent performance, SpecMini meets the testing and measurement requirements of the majority of RF signals. 2.Differentiation and advantage - Frequency range: 10MHz-4200MHz - DANL: -168 dBm @1GHz (Sensitivity set to High , normalized to 1Hz) - RBW: 10Hz-5MHz - Multi-screen: maximum 4 windows - Android operating system: touch screen operation, multitouch, easy-to-use. - Compact size (200mm×96mm×67mm) and light weight (1.25kg, including the battery) - 6 hours full operating time battery - Provide effective measurement guarantee during the building and maintenance of the transmission system 3.Target market State Radio Regulatory Commission, Operator, Research and Development, emergency communication, Armed Police, Army

Parameter

Fuction

Measurement

ACPR,CH Power ,OBW,N dB Down, Phase Noise

Trace

 3 Traces, TraceA, TraceB, TraceC

Trace Operation: Clear/ Write, Maxhold, Minhold, Average, View, Hide

Sweep

Continuous/Single

Trigger

Free

Video(Zero Span)

Marker

8 maximum, surport 1 reference with 7 Delta Markers

Peak Search, Marker Delta, Marker+, Marker-

Span

0Hz,100Hz to 4.2GHz

Freq Slip, Span Zoom(1-2-5-10Sequence,4GHz,4.2GHz)

Scale

Log/Lin

Log :0.1 to 1  0.1 step

   1 to 10  1 step

Unit

dBm, dBuV, dBV, W, mW, uW, pW, V, mV, uV

Sensitivity

Low, Medium, High

Detector

Positive, Negative, Sample, Average, RMS

Ref Level Offset

-50dB to 50dB

Multi-screen

4 Maximum

Frequency

Frequency Range

10MHz to 4.2GHz

Frequency Reference

Aging:±1ppm

Frequency readout accuracy

±( (readout frequency + 1GHz) x Frequency Reference + Frequency Span Accuracy x Span)

Frequency Span Accuracy

±1%

Sweep Time

1.1ms to 1600s

2.69ms to 1600s Zero Span

Resolution Bandwidth

RBW Range

10Hz to 5MHz,(1-2-3-5-10Sequence)

RBW Accuracy

RBW≥1MHz, ±10%   

RBW<1MHz, ±2%

Amplitude

Measurement  Range

DANL to +20dBm

Input Attenuator Range

0 to 30dB,1dB step

Max Safe input level

Sensitivity: low    +30dBm

Sensitivity: medium    0dBm

Sensitivity: High    -20dBm

Reference Level range

-140 dBm  to +20dBm

-190dBm  to +70dBm (Ref Level Offset :ON)

Amplitude Accuracy

ATT set to 0 dB , input signal -5 to -30 dBm, Detector set to Positive, Sensitivity set to Low, RBW auto-coupled, all other settings auto-coupled, 23±5℃.

Half hour warm-up required.

          ±1.5dB

RBW Switching uncertainty

±0.3dB

Input Attenuator uncertainty

±0.6dB

Reference Level Accuracy

RL≥-60dBm,±0.8dB

Display Average Noise Level(DANL)

Input terminated,  Detector set to Positive, Trace Average set to 1000, Span set to 50kHz,Reference level of -100 dBm, , all other settings auto-coupled, 23±5℃,

normalized to 1 Hz RBW

Sensitivity :LOW

1GHz,-131dBm/Hz(typical -133dBm/Hz)

Sensitivity:Medium

1GHz,-151dBm/Hz(typical -153dBm/Hz)

Sensitivity:High

1GHz,-168dBm/Hz(typical -169dBm/Hz)

Residual Responses

-75dBm

Input Related Responses

10M to 1.285G, 3.22G  to 3.7G                 <-70dBc

1.625G to 1.775G                              <-55dBc

1.285G to 1.625G, 1.775G to 2.35G, 2.71G to 3.22G  <-42dBc

3.7G to 4.2G                                   <-35dBc

2.35G to 2.71G                                <-25dBc

Second harmonic distortion

 1.6GHz   -70dBc

Third-Order Intercept (TOI)

(–10dBm tones ,1MHz apart, Sensitivity LOW, Reference Level –10 dBm

+15dBm

P1dB

+5dBm Norminal

Phase Noise

1GHz

-96dBc/Hz,@10kHz(typical -98dBc/Hz)

-118dBc/Hz,@1MHz(typical -120dBc/Hz)

General

System Setting

Preset, SaveRecall, ScreenCopy, Print, Language, Service, About

Connectors

RF In:N type, female, 50 Ω

USB:USB type C

Power interface: Slim Tip, DC20V

Display

5.5 inch, 1280*720p

OS

Android

Battery

Type: Li-on

Operation time:

option16 hours  norminal

option2:3.5 hours norminal

Charging time:

option1:2.5 hours norminal

option2:2 hours  norminal

Charging: 0°C to +35°C

Must be in power off state

Environmental

Operating Temperature: 0°C to 50°C

Storage Temperature: -20°C to 70°C

Size

200mm*99mm*67mm(7.8 in*3.9 in* 2.6in

Weight

1.25kg(2.8 lb) with battery option1

1.05kg (2.3lb) with battery option2

Technical specifications
This technical specifications include the influence of probability distribution, measurement uncertainty and environmental factors on the instrument performance. It guarantees the performance under the following conditions.
• The instrument is ON and warmed up for 30min.
• The instrument internal reference signal is applied.
Testing temperature is 23±5 ºC, unless other specific condition applied.
Typical value
Additional description does not cover all performance information of the product guarantee. Unless otherwise specified, the typical value refers to the indicator or technical specification with which more than 80% of products comply under 23 ± 5 ºC. The measurement uncertainty is excluded. SpecMini should be within the calibration period.
Nominal value
The nominal value refers to the characteristic description or design range. It is not tested or covered by the product. SpecMini should be within the calibration period.